With an equipment using Hopkinson pressure bar technique, and using a quartz gage pressure sensors,the high shock acceleration sensors were calibrated, and the calculating formula of the accelerometer sensitivity was developed. The experimental results show that the equipment designed to calibrate high shock acceleration sensors using a quartz gage pressure sensors can meet the calibration precision prescribed completely, and the whole calibration system has the advantage of simple structure, ease in library construction.